Chapter 3: Using BD FACSDiva Software 95
Verifying that the Configuration Matches the Sort Setup
To verify that the sheath pre ssure in the Sort Setup matches the sheath pressure in
the configuration:
1 Select File > Exit to exit CS&T.
2 Select Sort > Sort Setup, and select the Sort Setup that matches the nozzle
size listed in your configuration.
3 Select Cytometer > Sheath Pressure and verify that the pressure matches the
pressure in your configuration. Enter a new sheath pressure, if needed.
•To change a filter, click and drag a new
filter from the Filters list to the filter slot.
•To change a mirror, click and drag a new
mirror from the Mirrors list to the mirror
slot.
filter slot
mirror slot
Select the Sort Setup that
matches the nozzle size
listed in your configuration.