182 BD FACSAria II User’s Guide
2 Turn on the deflection plates.
Click the Voltage button ( ) in the Side Stream window. The voltage
warning light illuminates, indicating that the plates are charged.
NOTE Make sure the center stream image does not move after the plates
are turned on. Major movement of the center stream could indicate that the
plates or area around the plates needs cleaning. See External Cleaning on
page 193.
3 Click the Test Sort button and optimize the far left side stream, then turn
off Test Sort.
Adjust the far left slider for minimal deflection of the stream. The stream
should be deflected just enough to clear the hole in the splash shield. Do
not adjust the other sliders.
If you cannot see a stream image or the image is dim, adjust the micrometer
dial on the diode laser (Figure 1-20 on page 38) to better view the streams.
4 Select Sort > Home Device.
5 In the Device Setup dialog, select the collection device you are using and
click Go to Home.
Test Sort button