Chapter 7: Troubleshooting 241
No deflection or
insufficient deflection
Insufficient voltage Increase the side-stream voltages
using the slider controls.
•Increase the plate voltage.
Stream-charging wire is loose
or missing
Verify the stream-charging wire is
inserted all the way into the barb.
Salt bridge Turn off the stream. Remove the
nozzle. See Changing Nozzles on
page 214 for instructions.
Sorting paused because actual
Drop 1 value is out of range
Wait until the Sweet Spot adjusts
the amplitude to achieve the
Drop 1 target.
If this happens repeatedly during
sorting, there might be debris in
the nozzle or flow cell. See
Troubleshooting the Stream on
page 234 for suggestions.
Sorting Troubleshooting (continued)
Observation Possible Causes Recommended Solutions