Chapter 4: Running Samples 141
Figure 4-17 Deselecting Height parameter
Optimizing PMT Voltages
1 Right-click Cytometer Settings in the Browser, then select Application
Settings > Create Worksheet.
A second global sheet is added with the plots created according to your
selections in the parameters tab. (See Figure 4-18.) Use the gray boxes and
crosshairs to guide your optimization.
Figure 4-18 Optimizing PMT voltages
2 Load the unstained control tube onto the cytometer.
3 In the Cytometer window, optimize the settings for your application.
a Optimize the FSC and SSC voltages to place the population of interest
on scale.