192 BD FACSAria II User’s Guide
Figure 6-2 Connecting air and fluid lines for shutdown procedure
5 When prompted, install a tube containing 3 mL of 15% Contrad 70
cleaning solution the loading port, then click Done.
The cytometer loads the tube and continues the cleaning process. A
progress message appears, and then displays Done when the process is
complete.
air line
fluid line
ethanol shutdown
tank
sheath tank
sheath filter
air line
fluid line
ethanol shutdown
tank
sheath tank
disconnect
here
Normal operating configuration Fluid and air lines moved for shutdown
Przeglądanie stron 191