122 BD FACSAria II User’s Guide
Setting Up the Fluidics Cart
Check the fluid levels in the sheath tank and waste containers every time you use
the cytometer. This ensures that you will not run out of sheath fluid during an
experiment or have to service the containers during a sort. Fluidics level
indicators are shown in the Cytometer window in BD FACSDiva software. See
Figure 4-2 on page 115.
NOTE When the fluidics cart is in the correct position in the lab, lock the rolling
casters by turning the orange disk until the feet contact the floor. This helps
minimize any vibrations that could affect the stability of the stream.
If a fluidics container is running low, refill it with the fluid indicated on the
container label. For best results, fill the containers only with the fluids shown in
Table 4-2. For ordering information, see Reagents on page 268.
For instructions on emptying the waste, see page 127.
Table 4-2 Recommended fluids
Container Compatible Fluids Capacity
Sheath
a
a. Users should select solutions that are compatible with their specific samples and experiments.
BD FACSFlow™ sheath fluid (non-sterile)
1X phosphate-buffered saline (PBS), with or without
preservatives
one 10-L
stainless steel
container
Ethanol
Shutdown
70% solution diluted in DI or laboratory-grade water one 5-L stainless
steel container
Bleach 10% household bleach (0.5% sodium hypochlorite)
•BD FACSClean solution
5 L
DI Deionized water (Milli-Q or equivalent). Add 3 mL of
bleach per liter of DI water.
5 L
Ethanol 70% solution diluted in DI or laboratory-grade water 5 L