80 BD FACSAria II User’s Guide
Fluidics Shutdown
Fluidics shutdown removes sheath from the lines and replaces it with ethanol,
and cleans the cuvette with a cleaning solution. See Fluidics Shutdown on
page 190 for more details.
Change Sample Filter
Sample line filters can be installed to filter a sample before sorting. This selection
opens the Sample Filter Change wizard, which guides you through the process.
See Installing or Removing a Sample Line Filter on page 220 for the instructions.
Cleaning Modes
BD FACSDiva software contains pre-programmed cleaning modes that are
activated by choosing the corresponding menu command from the Cytometer >
Cleaning Modes menu. See Internal Cleaning on page 194 for more information.
Sheath Pressure
The sheath pressure determines how quickly particles pass through the laser
beam. Select Cytometer > Sheath Pressure to view the current sheath pressure and
change the pressure for custom sort setups.