Aria 8.0 Podręcznik Użytkownika Strona 102

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 101
102 BD FACSAria II User’s Guide
Table 3-2 Default Sort Setup values
Setting 70 micron 85 micron 100 micron 130 micron
Sheath Pressure 70 45 20 10
Amplitude 60 32 12 24
Frequency 87 47 30 12
Drop 1 150 150 150 150
Gap (upper limit) 6 (14) 7 (17) 10 (21) 12 (21)
Attenuation Off Off Off Off
Drop Delay 47.00 30.00 27.00 16.00
Far left voltage 100 100 80 60
Left voltage 40 35 30 20
Right voltage 40 35 30 20
Far right voltage 100 100 80 60
Plate voltage 4,500 4,000 2,500 2,000
2nd Drop 20 20 10 0
3rd Drop 10 10 5 0
4th Drop 0 0 0 0
Laser Delay (blue) 0.00 0.00 0.00 0.00
Laser Delay (red) -36.00 -54.00 -40.00 -82.00
Laser Delay (violet) 36.00 54.00 40.00 82.00
Area Scaling (blue) 1.40 1.20 1.40 0.75
Area Scaling (red) 1.00 0.75 1.00 0.50
Area Scaling (violet) 0.80 0.35 0.50 0.25
Window Extension 2.00 2.00 2.00 4.00