Figure 5-1 Turning on the deflection plates
Make sure the center stream image does not move after the plates are
turned on. Major movement of the center stream could indicate that the
plates or area around the plates needs cleaning.
4 Click Test Sort and optimize the side streams. See Figure 5-2.
Adjust the voltage sliders to view the required number of streams.
Figure 5-2 Turning on Test Sort
A 12,000-volt potential exists between the deflection plates when they are
on. Contact with the charged plates results in serious electrical shock. Do
not touch the deflection plates when the plate voltage is on. The plates
remain energized even when the sort block door is opened.