310 BD FACSAria II User’s Guide
3 Wait at least 30 minutes (115 V model) or 60 minutes (110 V model) to
allow the recirculating water reach the required temperature.
Maintenance
To maintain the recirc ulat ing water bath, refer to the documentation provided by
the manufacturer.
Tube Holders
Remove the tube holders when you are finished using them, and clean them
periodically before storage.
To detach t he recirc ulating water tubing, push in the orange collar as you pull the
tubing out of the port.
Clean the temperature control tube holders by wiping them down with an
appropriate cleaning fluid (ie, 70% ethanol, 5% bleach, deionized water). Dry
them with a lint-free cloth before storage.
Recirculating the Water Tubing
Inspect the tubing periodically for leaks, plugs, or contaminants. If needed,
remove the tubing and clean it with an appropriate cleaning solution, or replace
the tubing. Contact your BD Biosciences service representative for replacement
tubing.
When you detach the recirculating water tubing from the tube holder, any
fluid remaining in the tubing can leak into the sort collection chamber. To
ensure that fluid is aspirated from the sort collection chamber, make sure
the sort chamber aspirator pump is on (ie, do not turn off the instrument
main power) before you detach the tubing. Use caution when handling
tubing containing hot water.