Appendix D: Temperature Control Option 309
Starting Up the Water Bath
NOTE To ensure that the sample collection device is at the correct temperature,
start up the water bath at least 30 minutes (115 V model) or 60 minutes (110 V
model) before you start sorting.
1 Switch on the main power on the water bath control panel.
2 Use the up or down arrow keys to set the required temperature.
NOTE To achieve the requir ed sa mple temperature, you will need to set the
water bath temperature slightly higher or lower (see Table D-1). These
settings might need adjustment depending on the ambient temperature in
your laboratory. We recommend that you calibrate the water bath for your
operating environment.
Table D-1 Water bath settings for corresponding sample temperature
Required Sample
Temp (°C)
Water Bath Setting
42
37 37.5
42 43.2
main power
switch