148 BD FACSAria II User’s Guide
conjugates due to lot-to-lot variation. Refer to the BD FACSDiva Software
Reference Manual for more information about this feature.
A compensation specimen is added to the experiment, along with a stained
control tube for each compensation control. (Expand the specimen to view
all tubes.) Worksheets containing appropriate plots are added for each
compensation tube.
Calculating Compensation
The unstained control will be used to verify the settings for FSC, SSC, and FSC
threshold, and to gate the population of interest.
1 Install the unstained control tube onto the cytometer.
2 Expand the compensation specimen in the Browser.
3 Set the current tube pointer to the unstained control tube and click Load.
4 Verify that the population of interest is displayed appropriately on the FSC
vs SSC plot. Adjust if needed. (See Figure 4-22).
Since the application settings have already been optimized for your sample,
the cytometer settings should require little or no adjustment.