162 BD FACSAria II User’s Guide
Before you begin, you should be familiar with BD FACSAria II operation using
BD FACSDiva software. Review chapters 2 and 3, if needed.
Setting Up for Sorting
In general, do the following to set up for a sorting experiment. Each step is
explained in more detail in previous or subsequent sections.
1 Start up the cytometer and the computer.
See Cytometer Startup on page 114.
2 Install the appropriate size nozzle.
3 Select a new sort setup mode, if needed, and select an appropriate
cytometer configuration.
To c hang e your c urre nt s ort setu p mode, select a nozzle size from the Sort >
Sort Setup menu.
For more information, see Sort Setup on page 101.
4 Install a sample line filter, if needed.
See Installing or Removing a Sample Line Filter on page 220.
5 Check the laser delay for your sheath pressure and particle size.
To set the laser delay automatically using the Cytometer Setup and
Tracking module, see Checking Cytometer Performance on page 130.
To set the laser delay manually, see QC Using BD FACSDiva Software
on page 313.
6 Optimize cytometer settings for the sample to be sorted.
See Data Collection on page 144.