28 BD FACSAria II User’s Guide
Nozzle
The BD FACSAria II cytometer is provided with four nozzle sizes (70, 85, 100,
and 130 Pm) to accommodate a variety of particle sizes. The nozzle is keyed to a
fixed position at the lower end of the cuvette. Because the nozzle is below the
interrogation point, optical alignment is not affected when the nozzle is changed.
Figure 1-11 Nozzle
When installing or removing the nozzle, make sure the O-ring is in place before
proceeding. Look at the O-ring in the nozzle with a magnifier or under a
microscope to confirm it is positioned correctly.
Refer to these sections for more information on the nozzle:
•Changing Nozzles on page214
Cleaning the Nozzle on page 216
O-ring