232 BD FACSAria II User’s Guide
Removing or Installing the FSC ND Filter
For applications involving large particles where events appear off scale on the
FSC axis with a voltage of zero, keep the FSC ND filter in place to decrease the
FSC signal and keep the events on scale. For applications involving small
particles (eg, bacteria or platelets), you might need to remove the FSC ND filter
as follows.
1 Open the flow cell access door.
2 Locate and pull out the FSC ND filter (Figure 6-32).
The filter is installed at the left end of the FSC detector block, just to the
right of the flow cell. To remove the filter, loosen the set screw and pull the
filter out of the slot.
Figure 6-32 Removing the FSC ND filter
3 To reinstall the filter, slide it into the slot with the filter side down.
NOTE You can remove the nozzle holder to get better access to the ND
filter slot.
FSC ND filter
Set screw
Nozzle holder