220 BD FACSAria II User’s Guide
4 Insert the tubing into the union fitting and slowly tighten the nut until
secure. Do not over-tighten. Pull gently on the tubing to ensure it is secure.
Installing or Removing a Sample Line Filter
Sample filters can be installed on the end of the sample line to filter out large
particles from a sample. Pre-filtering the sample before beginning any sorting is
recommended.
Two sizes of sample line filters are included in the accessories kit, 35 micron (red)
and 50 micron (purple). The filters can be changed as often as required. The
sample filters are not intended for use with the 1 mL micro-tubes.
NOTE When a sample filter is installed, the sample flow rate can be slowed
down due to the effect of particles in the sample fluid clogging the filter.
1 Select Change Sample Filter from the Cytometer menu.
Make sure the stream is turned on before starting the process.
2 A wizard appears with the instructions. See Figure 6-21.
Figure 6-21 Sample line filter wizard