Chapter 6: Shutdown and Maintenance 217
Do not wipe the nozzle with anything, because it could leave fibers or other
contamination, or dislodge the O-ring.
6 Carefully reinsert the nozzle into the flow cell.
Push the nozzle gently all the way forward until it stops.
7 Turn the nozzle-locking lever clockwise to the 12:00 position.
8 Turn on the stream and make sure it flows through the nozzle properly.
9 Close the flow cell access door.
Note that after re-installing the nozzle, you might need to change the angle of the
sort block to re-center the stream in the aspirator. To do so, loosen the
adjustment screws on both sides of the deflection plates and rotate the sort block
(see Figure 1-12 on page 29). Tighten the screws when the stream is centered in
the aspirator. For further assistance, see Troubleshooting the Stream on page 234.