Chapter 7: Troubleshooting 239
Sorting Troubleshooting
Observation Possible Causes Recommended Solutions
Unstable breakoff while
Sweet Spot is engaged
Residual ethanol in system Allow the system to run until the
breakoff stabilizes.
Target Drop 1 value is out of
range for drop spacing
Use an actual Drop 1 value for the
target. Remember to repeat the
drop delay setup each time the
target value is changed.
Nozzle is clogged or
inserted improperly
Turn off the stream. Remove the
nozzle and clean it. Re-insert the
nozzle and slide the nozzle in until
it stops, then close the locking
lever.
Dirty strobe lens or upper
camera window
Clean the lens and the window as
described in Cleaning the Camera
Windows on page 225.
Air in sheath filter Purge the sheath filter.
Debris in flow cell or nozzle Remove the nozzle and run the
stream with no nozzle in place for
approximately 10 seconds. (Click
the Stream control on, and then
off.) Sonicate the nozzle and
re-install it.
Center stream image is
dim or not visible in the
Side Stream window
Camera window is dirty Clean the lower camera window.
See Cleaning the Camera
Windows on page 225.
Stream is not intercepting
the diode laser
Adjust the micrometer dial on the
laser to make the stream intercept
as bright as possible. See
Figure 1-20 on page 38.