124 BD FACSAria II User’s Guide
5 Fill the tank with sheath fluid up to the upper weld line on the inside of the
tank. See Figure 4-5.
NOTE Do not overfill the sheath tank because this can cause incorrect
sample flow rates.
Figure 4-5 Sheath tank fill level
6 Replace the cover and tighten the knob.
Make sure the large O-ring on the inside lip of the cover is seated correctly
and has not slipped out of position. The tank can leak if the cover is not
secured properly.
7 If you removed the sheath tank from the fluidics cart to refill it, always
place the tank back in its original position on the cart. See Figure 4-6.
The flow rate is calibrated with the sheath tank on the fluidics cart. If the
location or elevation of the sheath tank is changed, it could affect the flow
rate calibration.
Figure 4-6 Sheath tank location label
upper weld line to
indicate fill level
ATTENTION
SHEATH TANK
(
W
)
to be
LOCATED HERE
during instrument operation
to maintain calibration