Chapter 6: Shutdown and Maintenance 193
6 Click OK when you see a message informing you the system can be turned
off.
7 Turn off the cytometer main power.
8 Exit BD FACSDiva software and shut down the computer.
External Cleaning
To keep the system free from sa lt bu ildup, wipe down all cytometer surfaces that
have been exposed to sheath fluid. Clean surfaces with a cloth dampened with a
10% bleach solution, followed by DI water.
The following surfaces should be inspected and cleaned when necessary:
•Inside the sort chamber
•Deflection plates
•Sample loading port
•Collection devices
All cytometer surfaces that come in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when cleaning
cytometer surfaces. Wear suitable protective clothing and gloves.
To prevent shock, turn off the plate vo ltage before cleaning on or around
the deflection plates. To prevent arcing (sparking), make sure the plates are
completely dry before you turn the plate voltage back on.