202 BD FACSAria II User’s Guide
Changing Fluid Filters
We recommend changing the fluid filters every 6 months. Spare filters are
included with the accessory kit.
Figure 6-7 Fluid filters
1 Remove the filter by pressing the tabs on each quick-disconnect coupling.
(See Figure 6-7.)
2 Install the new filter and connect the quick-disconnect couplings.
Write the current date on the filter so you will know when to replace it.
3 Open the bleeder valve on top of the filter a small amount and leave it open
until fluid seeps out through the valve.
4 Close the valve.
5 Wipe up any excess fluid that might have dripped onto the fluidics cart.
bleeder
valve
quick-disconnect
coupling
quick-disconnect
coupling