126 BD FACSAria II User’s Guide
6 Replace the cover and tighten the knob.
Make sure the large O-ring on the inside of the cover is seated correctly and
has not slipped out of position. The tank can leak if the cover is not secured
properly.
7 If you are going to perform fluidics shutdown, connect the air line.
Refilling the Plastic Containers
You can refill the plastic containers directly on the fluidics cart without detaching
any lines, or you can remove the container for refilling. Note that during
operation, you can add fluid to a container through the large cap without any
interruption to your experiment, but if you detach any lines, you will need to
prime the system.
1 (Optional) Disconnect the container’s sensor and quick-release connector if
you need to move the container. (See Figure 4-8.)
If you do not need to move the container, skip to step 2.
Figure 4-8 Plastic fluidics container lines
2 Remove the large-sized cap from the container (See Figure 4-8.)
3 Fill the container with the fluid indicated on the container label.
quick-release
sensor
connector
large-sized
cap