136 BD FACSAria II User’s Guide
Application Settings
Application settings are associated with a cytometer configuration and include
the parameters needed for the application, area scaling values, PMT voltages, and
threshold values, but not compensation. Each time a performance check is run
for a configuration, the application settings associated with that configuration
are updated to the latest run.
Using application settings provides an easy, consistent, and reproducible way to
reuse cytometer settings for your commonly used applications. Refer to the
Cytometer Setup and Tracking Application Guide for information on running a
baseline and for other details on using the CS&T application.
Before creating application settings you must:
•Perform the cytometer startup procedure described in Cytometer Startup
on page 114.
•Run a performance check for the cytometer configuration that will be used
for the application.
Creating Application Settings
This section describes how to create and save application settings for a multi-
color immunophenotyping sample.
Setting Up the Workspace
1 Create a new folder.
2 Create a new experiment.
3 Select the Cytometer Settings in the Browser.
4 In the Inspector window, click the Parameters tab and delete unneeded
parameters.