Chapter 7: Troubleshooting 237
Leaking or spraying
around nozzle
Defective or damaged nozzle
O-ring
Replace the O-ring.
Nozzle inserted improperly Turn off the stream. Remove the
nozzle. See Changing Nozzles on
page 214 for instructions.
Extra O-ring is blocking the
nozzle
Remove the nozzle and use a
cotton swab to clear out the
cuvette.
Drop breakoff is too
long
Bubbles in flow cell Open the flow cell access door
and check for bubbles in the flow
cell. If they are visible, turn off the
stream, wait a few seconds, and
turn on the stream again.
Attenuation is on Turn off attenuation.
Amplitude is too low Increase the amplitude until you
can see drops. If you need a very
high amplitude (>70 volts) to see
drops, you might have air bubbles
in the flow cell.
Nozzle inserted improperly Turn off the stream. Remove the
nozzle and ensure that the O-ring
is in place. Re-insert the nozzle
and slide the nozzle in until it
stops, then close the locking lever.
Problems using Auto
Delay feature
Stream is not stable Make sure stream is stable before
starting to run auto delay.
Diode laser does not fully
intercept the sorting streams
Adjust the diode laser to intercept
streams in the middle, producing
biggest and brightest spots in the
left and center images.
Event rate is too low or too
high
Adjust flow rate to increase or
decrease event rate.
Troubleshooting the Stream (continued)
Observation Possible Causes Recommended Solutions