Chapter 4: Running Samples 125
8 Connect the air line. The system is ready to run again.
Refilling the Ethanol Shutdown Tank
The fluid level in the stainless steel ethanol shutdown tank should be checked
before starting the fluidics shutdown procedure and refilled when low. This tank
does not have a sensor, so it must be checked manually.
Figure 4-7 Ethanol shutdown tank connectors
1 Turn off the stream.
2 If it is connected, disconnect the air line on the ethanol shutdown tank.
3 Pull up on the ring of the pressure relief valve to release pressure from the
tank.
Make sure the
tank is fully vented.
4 Unscrew the tank cover knob and remove the cover.
5 Fill the tank with
ethanol up to the weld line. See Figure 4-7.
cover knob
air connector
pressure
relief valve
fluid connector
weld line