82 BD FACSAria II User’s Guide
Sample Temperature
Use the Sample Temperature command to set the temperature inside the sample
injection chamber. You can select from one of the specified values, or select Off to
turn off temperature control (Figure 3-2).
Figure 3-2 Setting the sample temperature
The cooling/heating unit is designed to maintain the temperature of a sample
tube. It is not designed to cool or heat the sample. It takes approximately 45
minutes to reach the required temperature inside the chamber, during which time
the chamber must be kept closed.
To c ool or heat the chamb er, in stall a s ampl e tube and click Load. Click S top
Acquiring to keep the chamber closed and stop running the sample. To maintain
the temperature, do not leave the chamber open for extended periods while
changing sample tubes.
The sample temperature retains the last setting after startup. For example, if it
was set to 20 C the last time the system was used, then it will return to that
setting the next time the system is started up.
cooling/heating
unit for sample
injection chamber